<?xml version="1.0" encoding="utf-8"?>
<script>
  <name>Enable bridge - external reference w/ AC excitation 60 SPS Sinc3</name>
  <description />
  <command_list>
    <command>
      <command_string>RESET</command_string>
      <description>Reset device</description>
    </command>
    <command>
      <command_string>WREG 02 32</command_string>
      <description>SINC3, 60 SPS</description>
    </command>
    <command>
      <command_string>WREG 03 61</command_string>
      <description>Enable 4-wire AC excitation</description>
    </command>
    <command>
      <command_string>WREG 04 C0</command_string>
      <description>Enable GPIO Outputs on AIN4/5</description>
    </command>
    <command>
      <command_string>WREG 05 6C</command_string>
      <description>Enable STATUS/CRC bytes</description>
    </command>
    <command>
      <command_string>WREG 06 0A</command_string>
      <description>External AIN0/1 reference selected</description>
    </command>
    <command>
      <command_string>WREG 10 07</command_string>
      <description>PGA enabled, gain of 128 V/V</description>
    </command>
    <command>
      <command_string>WREG 11 34</command_string>
      <description>Select AIN2/3 as analog inputs</description>
    </command>
    <command>
      <command_string>DELAY 2000</command_string>
      <description>Delay for settling...</description>
    </command>
    <command>
      <command_string>SYOCAL </command_string>
      <description>Run system offset calibration</description>
    </command>
    <command>
      <command_string>DELAY 10</command_string>
      <description>Delay for calibration</description>
    </command>
    <command>
      <command_string>CLEAR</command_string>
      <description>Clear STATUS register</description>
    </command>
    <command>
      <command_string>REGMAP</command_string>
      <description>Readback register settings</description>
    </command>
    <command>
      <command_string>STATUS</command_string>
      <description>Check for faults</description>
    </command>
    <command>
      <command_string>DELAY 10</command_string>
      <description>Delay for settling</description>
    </command>
    <command>
      <command_string>COLLECT 300</command_string>
      <description>Collect 300 points of data</description>
    </command>
  </command_list>
</script>