Integration TIDA-00096 and TIPD111?

Prodigy 50 points

Replies: 3

Views: 699

Hello Everybody,

I want to know is it possible to integrate  TIDA-00096 ECG monitor and  TIPD111 hardware pace detector.

Thank you in advance.

  • Hi Alexander,

    I've moved this post from the Medical Forum to the Precision Data Converter Forum, where the TI design TI-00096 is supported. I'm sure you'll find an answer here.

    Regards

    Praveen

    TI makes no warranties and assumes no liability for applications assistance or customer product design. You are fully responsible for all design decisions and engineering with regard to your products, including decisions relating to application of TI products. By providing technical information, TI does not intend to offer or provide engineering services or advice concerning your designs.

  • In reply to Praveen Aroul:

    Hello Alexander,

    TIPD111 requires a ECG signal to scan for pace. ADS1293, the AFE on TIDA-00096, has the ability to output the multiplexed ECG channels to either the RLDIN or WCT pins, but I do not believe those nets are brought to headers on TIDA-00096. Perhaps someone more familiar with the ADS1293 can comment.

    Brian
  • In reply to Brian Pisani:

    Hi Alexander,

    The Pace signal is available on the WCT pin of the ADS1293. It is not brought out to a connector on the TIDA-00096 design, but is brought to a via that you might be able to solder to.

    Mike